Please use this identifier to cite or link to this item: http://repository.ucc.edu.co/handle/ucc/1290
Title: Quality control of the breast ca treatments on HDR brachytherapy with TLD-100
Author: Torres Hoyos, Francisco José
De la Espriella Velez, Nicolas
Sánchez Caraballo, Alvaro Antonio
Email autor: francisco.torresh@campusucc.edu.co
Issue Date: 2014
Keywords: TLD-100;Anthropomorphic phantom;Brachytherapy HDR;Breast
Abstract: An anthropomorphic Phantom, a female trunk, was built with a natural bone structure and experimental material coated, glycerin and waterbased material called JJT to build soft tissue equivalent to the muscle of human tissue, and a polymer (styrofoam) to build the lung as critical organ to simulate the treatment of breast cancer, with high dose rate brachytherapy (HDR) and sources of Ir-192. The treatments were planned and calculated for the critical organ: Lung, and injury of 2 cm in diameter in breast with MicroSelectron HDR system and the software Plato Brachytherapy V 14.1 of the Nucletron (Netherlands) which uses the standard protocol of radiotherapy for brachytherapy treatments. The dose experimentally measured with dosimeters TLD-100 LIF: Mg; Ti, which were previously calibrated, were placed in the same positions and bodies mentioned above, with less than 5% uncertainty. The reading dosimeters was carried out in a Harshaw TLD 4500.The results obtained for calculated treatments, using the standard simulator, and the experimental with TLD-100, show a high concordance, as they are on average a § 1.1% making process becomes in a quality control of this type of treatments.
Table Of Contents: Resumen Introducción Metodología Resultados y discusión Conclusiones Agradecimientos Referencias
Program: Ingeniería de Sistemas
Headquarters: Montería
Publisher: Universidad Cooperativa de Colombia, Facultad de Ingenierías, Programa de Ingeniería de Sistemas, Montería, Colombia, 00000
Type: Artículo
CC Licence: Licencia CC
Citation: Torres Hoyos, F. J., De la Espriella Vélez, N., & Sánchez Caraballo, A. A., (2014). Quality control of the breast ca treatments on HDR brachytherapy with TLD-100, Revista Mexicana de Física 60, 409–413.
Resource reference: https://rmf.smf.mx/pdf/rmf/60/6/60_6_409.pdf
Appears in Collections:Ingeniería de Sistemas

Files in This Item:
File Description SizeFormat 
Quality Control of the breast ca treatments.pdfarticulo principal quality control506.62 kBAdobe PDFThumbnail
View/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.